Electromyography (facial or other)

What does this channel measure?

This channel measures muscular activity of specific facial muscles (e.g. m. corrugator or zygomaticus) or general tension related muscles (e.g. neck). High activity of the muscle is reflected in rapidly oscillating signals of large amplitudes. The raw signal is highpass-filtered at 28 Hz, rectified, and smoothed with a 16 Hz lowpass filter or moving average. The oscillating signal thus becomes a smoother curve with only positive values, that can directly be interpreted in terms of amount of muscle activity.

To change filter settings, you can edit the 'anslabdef.m'-file in your study folder and manually change the filter values. Alternatively you may change the respective values through the ANSLAB options dialog. They are read in before loading data and will be applied to the loaded data.

[Top]

Editing of EMG Data

This signal hardly requires any editing. Moreover, artifacts are difficult to separate from normal activity. You can however exlude outliers if you are sure about artifactual epochs and set them to missing values.